MT-1000 X-Y Translation
System
|

MT-1000 X-Y Translation System
X-Y TRANSLATION SYSTEM FOR FIXED-STAGE UPRIGHT MICROSCOPES
Sutter Instrument Company is pleased to introduce an innovative new
design for use in fixed-stage, moving optical pathway microscopy. The
modular design of the MT-1000 features free-standing columns that
provide rock solid mounting for the tissue recording chamber and
multiple micromanipulators, such as the MP-285 Motorized
Micromanipulator. The columns are positioned around a movable table that
provides X-Y translation of the optical pathway in order to view
multiple locations within the chamber at high power.
The column-mounted tissue recording chamber is relatively compact and
serves as an alternative to bulky stages. The modular columns allow easy
access to the microscope and provide a very flexible design platform for
customizing your rig.
At the core of the system is the X-Y translation table. It is
designed to allow movement of a microscope's optics with respect to the
fixed stage. Such an arrangement is currently in use by
electrophysiologists recording from tissue slice preparations. The
translator allows visualization and/or imaging of multiple tissue
locations while maintaining multiple stable recordings from the
preparation. Such a system is also useful for experiments on cells in
culture where one wishes to monitor several cells not in the same field
of view, e.g. recording from pre and post synaptic neurons in culture
and/or imaging one cell while recording from or stimulating another.
Translation tables are available for the Olympus BX-50, Nikon E600FN,
and Zeiss AxioskopFS. A generic rectangular table (295mm x 340mm) which
can accommodate most standard, upright microscopes as well as a number
of inverted microscopes is also available.
The recording chamber and micromanipulators are mounted on
independent columns firmly affixed to the translator baseplate. The
manipulator columns are identical to the popular MT-75 gantry style
stands. The recording chamber fits into a dovetail slide on a similar
column. While the chamber column shown opposite conforms to a Warner
Instruments Series 20 design, other chambers can be accommodated. A
dovetail slide with a generic opening is available for users who wish to
develop custom chambers.
Specifications
| (MT-500 X-Y Translator) |
|
| Baseplate Dimensions |
20in x 24in x 2in
500mm x 600mm x 50mm |
| Maximum Travel |
1in
25mm |
| Resolution |
0.0002in
5µm |
| Maximum Load |
110lbs
50kg |
| Weight |
35 lbs
16kg |

Back to top
Site by Rob Jones.
Copyright © 1999 Linton Instrumentation . All rights reserved.
Revised: August 11, 2000
. |