Capnometer for End Tidal CO2
|
|

|
NEW! MRI compatible version available with
reduced magnetics, enquire for
details.
|

Capstar 100 End-Tidal CO2 Analyser.
CapStar-100 End-Tidal CO2
Analyser. This fast-response infrared analyzer features excellent accuracy and
stability, and can be used with small subjects.
The Capstar-100 Carbon Dioxide Analyser provides continuous, accurate measurement of
expired CO2 over the range of 0-10%. It's rapid response time, moderate sample flow
requirements, and long term stability make it ideal for routine respiratory gas measurement
with any size subject.
The heart of the instrument is a miniature infrared optical analyser that features
exceptional stability, low maintenance, and a linear output voltage for external
recording. The heated temperature-controlled, measurement cell provides protection
from water condensation and an included moisture trap can be used to capture condensed
water in the sample tubing.
The remote sensor-pod, attached to the main unit by a 1.5m cable, contains the
measurement cell, which can be removed for inspection or cleaning.
The built-in digital display shows CO2 concentration in % or mmHg, and allows
breath-by-breath peak end-tidal reading or continuous rapid response to changing CO2.
Calibration is readily performed with a single calibrated gas and room air.
The Capstar-100 can be used in conjunction with any ventilator or simply to monitor
unassisted subjects.
Specifications:
Analyser Type
|
Single Beam Infrared
|
Sample Cell Materials
|
Sapphire and Stainless Steel
* Reduced Magnetics Version now available
|
Analogue Output Level
|
100mV/% CO2 (0-1V)
|
Measurement Range
|
0-10% CO2
|
Linearity
|
0.1% CO2
|
Repeatability
|
0.1% CO2
|
Zero Stability
|
0.1% (8 Hours), 0.2% (24 Hours)
|
Operating Temperature Range
|
5-40degC
|
Sample Rate Flowmeter
|
10-100ml/min
|
See Also:
Gemini Respiration 02, CO2 and Rate Monitor.