Ultra-Miniature Pressure-Volume Catheter
The Scisense Pressure-Volume system brings an advanced solution to
the problem of characterizing the cardiac phenotype of transgenics. Our
integrated catheter for simultaneously measuring LV pressure and volume
is the smallest such catheter in the world and creates
the least possible disruption in small hearts.
Only 1.2 French in size, our P-V catheter is the ultimate in the
assessment of hemodynamic parameters and cardiac function, allowing
continuous measurement of LV pressure and volume. In addition, our
sensors are small enough to slip through aortic bands or the lumens of
other devices, and no de-bubbling or flushing is required.
The P-V catheter includes our high fidelity pressure sensor (.39 mm
diameter) and four electrode rings (.37 mm diameter, .25 mm width). The
similar sizes of these components allow for a smooth catheter tip
without any bulges that can otherwise make insertion difficult.
For <100g subjects, the inner electrode segment length is 4.5 mm (center
to center). For larger subjects (100-500g), the inner segment length is
9.0 mm. For both types of catheter the outside electrode rings are
spaced 0.5 mm from the inner rings.
All models have the pressure sensor side mounted between the inner
electrodes. The associated pressure control box for the catheter
contains the signal acquisition and amplification circuitry, providing
an analog pressure signal through a standard BNC connection.
Scisense pressure-volume catheters are warranted for six months from
date of purchase.
Pressure-Volume Sensor Specification:
Measurement
Range
|
-50 to + 300 mmHg
(temperature range: 32-42°C) |
Sensitivity
|
10µV/V/mmHg or higher |
Frequency Response
|
0 to >10 kHz |
Gauge Excitation
|
Constant current |
Stability
|
Zero offset drift < 6 mmHg/12 hours * |
Signal Gain Factor
|
200 |
* after 15 minute pre-soak