You are here:   Products
  |  Login

Categories

Minimize

All Products

Enquiry List

Minimize

Your Enquiry List is empty

Products

Minimize
BAT10

Physitemp BAT-10 Multipurpose Thermometer

The BAT-10 is the most versatile thermometer Physitemp have ever designed.
Code
BAT-10

Physitemp BAT-10 Multipurpose Thermometer

The BAT-10 is the most versatile thermometer Physitemp have ever designed. Wide temperature range and extensive choice of probes can be used in applications as diverse as cryogenic measurements for blood banking and cryosurgery, skin temperature measurements in exercise experiments, liquid measurements in spectrophotometer cuvettes, melting points of plastics and all types of animal and insect temperature measurements. BAT-10 accuracy is NIST traceable and in each temperature range accuracy is the same as the resolution. For instance, in the 0.1°C range, accuracy is 0.1°C (± 1 least significant digit). Few digital thermometers will do this. In most instruments there is a difference of several tenths between specified accuracy and resolution. At physiological temperatures, where minute changes can be critical, the differential range allows readings to 0.01°C with 0.01°C accuracy.

Features:

  • Multiple inputs
  • Three ranges
  • NIST traceable accuracy
  • Resolution and accuracy as high as 0.01° C
High Accuracy in all Temperature Ranges

Range Resolution Accuracy
-100 to +200° C 0.1° C 0.1° C (+ 1 digit)
-200 to +400° C 1° C 1° C (+ 1 digit)
Differential 0.01° C 0.01° C (+ 1 digit)






Specifications

Temperature Range and Resolution: -200 ° C to +400 C, 1 ° C resolution

- l00°C to + 199.9°C, 0.1°C resolution



Differential Temperature Range: - 19.99° to + 19.99°. Linearization

centered at 40°C. 0.01°C resolution.



Repeatability: + one least significant digit
Calibration: Conforms to National Institute of Standards and Technology tables (Monograph 125)
Sensors: Any Physitemp type T thermocouple
Ambient OperatingRange: 15 - 45°C
Readout: Liquid crystal, 3 1/2 digits
Batteries: BAT-10 -- 4 alkaline "C" cells
(battery life, 1000 hours)

BAT-10R (rechargeable unit) - 4 Ni-Cads

Analog Output: Non-linearized set at 1.6V corresponding to temperature of 401°C