Biopac NIBP200A - Tail Cuff Non-Invasive BP Amplifier

The NIBP200A small animal tail noninvasive blood pressure amplifier incorporates a built-in pump that automatically inflates the blood pressure cuff to occlude the vessel in the tail of a rat or similar small animal.
Code
NIBP200A

Biopac NIBP200A - Tail Cuff Non-Invasive BP Amplifier

the blood pressure cuff to occlude the vessel in the tail of a rat or similar small animal. Once the pump reaches the inflation point it slowly deflates the cuff, providing a linear drop in pressure. A single pushbutton controls both the inflation and deflation cycles, making the system very operator friendly.

The cuff assembly attaches to the amplifier via a Luer lock connector and the Piezo pulse sensor attaches via a BNC connector. The amplifier unit also has a gain adjustment control to amplify or attenuate the pulse signal. The amplifier provides two analog outputs for pressure and pulse waveforms.

Each system comes with one cuff assembly, pulse sensor, restrainer, and interface cables (pulse and pressure). The interface cables provide simple connection to your BIOPAC or third-party A/D hardware. An optional tail heater is available.

The NIBP200A includes:
1 - amplifier unit (switching power supply works for 110 V or  220 V)
1 - cuff/sensor (size 11 mm is included, but can be specified as 9.5 mm or 13 mm when ordering)
1 - restrainer (size medium is included, but can be specified as small or large when ordering)
1 - Pulse cable to MP System
1 - Pressure cable to MP System

Optional accessories:
Tail Heating Unit TAILHEATB (220 V 50 Hz)
MRI-compatible sensors (contact us for more details)

Setup and calibration details are in the Hardware Guide (under Help in the BIOPAC software or online under Manuals).

NIBP200A - Tail Cuff Non-Invasive BP Amplifier Specifications:


Cut-off Pressure Range: 0-100 mmHg
0-150 mmHg
0-200 mmHg
0-250 mmHg
0-300 mmHg
Pressure Accuracy 300 mmHg Full Scale 1%
Pressure Sensitivity 0.1 mmHg
Pressure signal output: 300 mmHg/3 Volt DC
Pulse Gain: 5% to 100% (adjustable)
Pulse signal Output: 0- 4 Volt DC
CH 2 Pulse Display: Pulse intensity is displayed on CH 2, derived from plethysmographic measures. The tail sensor detects blood flow and pulse intensity is increased or decreased, depending on the flow ratio.
Analog output connectors: BNC
Power Supply: 12 Volt 1 Amp - External