Lambda DG-4 Ultra High
Speed Wavelength Switcher
|

Lambda DG-4 Ultra High Speed Wavelength Switcher
DG-4 ULTRA HIGH SPEED WAVELENGTH SWITCHER
The Lambda DG-4 offers unprecedented speed and versatility for
experiments requiring rapid light wavelength switching. It offers all
the advantages of interference filter- based systems, yet eliminates the
temporal constraints imposed by filter switching devices. Switching
between any two wavelengths is achieved in less than the 1.2msec
vertical retrace period of a video signal, allowing you to perform
real-time video imaging. For dual wavelength ratio imaging studies, the Lambda
DG-4 enhances your ability to follow fast changes in ion
concentrations by acquiring a ratio pair in two consecutive video
frames.
Narrow band-pass systems, such as single cavity interference filters,
grating monochrometers, and A.O. modulators, will, by their nature, pass
harmonics of the desired wavelength. With variable wavelength devices,
such as gratings, it is not always possible to obtain sufficient
blocking of out-of-band wavelengths. Modern interference filters, as
used in the Lambda DG-4, have integral blocking characteristics
1000 times better than typical monochrometer systems. In addition, it is
difficult to adjust the intensity of one wavelength relative to another
in a traditional system with a single optical path and a variable
wavelength device. The dual galvanometer design of the Lambda DG-4
eliminates this problem.
How it works:
This unique optical design of the Lambda DG-4 is based on dual
scanning galvanometers. One galvanometer is placed at the focus of the
arc lamp and is used to reflect the lamp output through one of four
optical channels that may contain an interference filter. After the
light passes through a filter, it is sent to a second scanning
galvanometer mirror that directs the light to a light guide which
provides spatial uniformity. The light guide can be coupled to the
illumination port of an instrument (e.g. epi-illumination port of a
microscope). By controlling the relative orientation of the two scanning
mirrors, intensity of the output can be modulated. Thus this system can
provide narrow band excitation at selected wavelengths over a range of
intensities or can rapidly turn off the light source. Dwell time at any
wavelength is arbitrarily set by the user. Transitions are achieved in
less than 1.2msec.
Back to top
Features
Complete system for wavelength switching
- A built-in 175 Watt ozone-free xenon lamp makes the Lambda DG-4 a
complete excitation system, and eliminates problems associated with
device integration.
- Up to four standard interference filters (25mm diameter, 9mm
thick) and neutral density filters can easily be installed in the
system.
- The light guide output from the Lambda DG-4 provides the
additional benefit of vibration isolation from your microscope .
Integral shuttering and filtering
- The Lambda DG-4 incorporates a high speed shutter with open/close
times of 500us. The shutter can be enabled between filter
transitions to prevent light transmission through intermediate
filters.
Integral neutral density filtering
- Neutral density filtering is achieved under program control by
offsetting the output galvanometer such that light is not centered
on the fiber optic light guide. Up to 15 filters can be defined with
this method. Due to the scrambling effect of the light guide, the
output still has excellent uniformity.
- Direct insertion of neutral density optical filters is also
acceptable in the filter holders at any of the four optical
channels. A final optical filter can also be placed in the exiting
light path which will reduce the light output from all 4 optical
channels.
Two outputs for monitoring filter position
- A 4 bit TTL signal transmits the current optical channel (filter)
position. A digital-to-analog converter (DAC) output produces a
voltage showing which filter is in use.
Methods of control
- Direct Computer Control via Parallel or Serial Interface
When operated in this mode, the Lambda DG-4 control commands are a
subset of our Lambda 10-2 controller and will operate with software
written for the Lambda 10-2. (to take full advantage of the additional
features of the Lambda DG-4, we recommend checking with your software
vendor for the appropriate driver)
- Strobe-Pulsed Ring Buffer Control
Up to 32 filter sequences can be downloaded into a ring buffer via
keypad or computer. The system will change to the next filter in the
buffer on a strobe pulse. Each subsequent pulse will cause a move to
the next filter in the buffer. After executing the last filter change
in the string the system resets to the first filter and continues.
A video sync pulse can be used to initiate the filter change for
each sequence stored in the ring buffer. Stand-alone operation of the
Lambda DG-4 is achieved in this mode using only a video sync input.
- Video Sync with Strobe Low
The system will implement a filter change on the video sync pulse,
but only if the strobe line is held low. This allows a computer to
override the sync pulse.
Back to top
Specifications
| Output Range |
300nm to 700nm |
| Output Power |
7.5mWatts @340nm
8.2mWatts @380nm (Measured at light guide output) |
| Lamp Type |
175 Watt ozone free xenon, pre-aligned |
| Lamp Life |
1,000 hours
(All bulbs warrantied for 500 hours. Longer life depends on your
application. Expected life is 1000 hours.) |
| Power Consumption |
350 Watts |
| Filter Diameter |
1in
25mm |
| Dimensions |
10in x 10in x 19in
25cm x 25cm x 48cm |
| Light Guide |
2 meters long
3mm wide |
| Weight |
45lbs
20kg |
| Electrical |
115/230 Volts
50/60 Hertz power line |
Back to top
Ordering Information
BASIC SYSTEM
DG-4/OF Includes Lambda DG-4 unit, pre-aligned 175 Watt ozone
free xenon bulb ,4 slide in filter holders, 1 neutral density filter
holder, serial and parallel cables, liquid light guide, and manual
DG-4/FS Same as DG-4 but with 175W full spectrum bulb (note:
full spectrum bulb produces ozone, and will require additional
ventilation of the lamp exhaust. Contact our technical support staff for
details)
ACCESSORIES
O661176* Ozone free 175 Watt xenon bulb (attenuated output
below 350nm, 80% at 340nm)
O661175* Full spectrum 175 Watt xenon bulb
DG-IF Interference filter holder
DG-ND Neutral density filter holder
X100160 Retaining ring
X100150 Filter spacer
* Please note that the above bulbs do not include the outer
blue housing. If you need or want an extra housing, please contact
us.
Mounting Adapters for Nikon, Zeiss, Leica and Olympus microscopes are
available. Please call Sutter Instrument for pricing and further
information. Also, please see the adapter list at the end of this
section.
Back to top